Abstract
As a key parameter of microwave dielectric ceramics, the temperature coefficient of resonant frequency (τf) determines the temperature stability of microwave dielectric ceramics and relevant microwave devices. The τf of a microwave dielectric ceramic is typically calculated from the measured resonant frequencies of the system, consisting of the ceramic sample plus other dielectric elements, at two temperatures ∼60 – 70 °C apart. This measurement process assumes that τf is a constant; however, it is generally temperature-dependent based on both theoretical and experimental results. Furthermore, the as-measured τf is actually the τf of the dielectric resonator system, which is different from that of the sample itself. In this paper, these two overlooked problems during τf measurement are discussed in detail, and suggestions are given to avoid their possible misleading consequences and thus ensure the reliability of τf measurements.
| Original language | English |
|---|---|
| Article number | 112920 |
| Journal | Materials Research Bulletin |
| Volume | 178 |
| DOIs | |
| State | Published - Oct 2024 |
Keywords
- Measurement
- Microwave dielectric ceramics
- Temperature coefficient of resonant frequency
Fingerprint
Dive into the research topics of 'On the τf measurement of microwave dielectric ceramics'. Together they form a unique fingerprint.Cite this